DataShuttle

Specifications

DataShuttle/USB

General

  • Isolation: 500 VDC from PC common
  • Power Requirements: Powered from USB, or from external +6 to +16 VDC (AC supply included)
  • Environmental: 0° to 50°C, 0-95% RH (noncondensing)
  • Dimensions: 182 mm W x 92 mm D x 45 mm H (7.1” x 3.6” x 1.8”)

Analogue Specifications

  • Each channel is configurable for single-ended or differential; volts or thermocouple inputs
  • DataShuttle/USB 55: 10 single-ended, 5 differential; volts or TC channels
  • DataShuttle/USB 56: 20 single-ended, 10 differential; volts or TC channels
  • Input Voltage Range: Sixteen software-programmable ranges, from ±20V full scale to ±30 mV full scale
  • Thermocouple Type: J, K, T, E, R, S, B, N14G, & N28G
  • Thermocouple Accuracy: ±0.5°C
  • Over-Voltage Protection: ±45V relative to analogue common
  • AC Common Mode Rejection: >120 dB @ 50/60 Hz
  • Channel-to-Channel Crosstalk: <-110 dB (DC to 100
  • Hz; up to 10K Ohm source resistance)
  • Accuracy: 0.01% of reading, +0.002% of range (exclusive of noise)
  • Input Resistance: 10M Ohm (SE), 20M Ohm (DE)
  • Bias Current: <1nA (0° to 35°C)
    Cold-Junction Accuracy: ±0.5°C (15° to 35°C)
  • Measurement Speed: Each channel can have independent measurement duration and resolution; channels can be programmed to be scanned in any order
  • Resolution/Speed (10 channel scan): 22 bit, 3 samples/s; 21 bit, 8 samples/s; 19 bit, 25 samples/s; 17 bit, 48 samples/s; 15 bit, 80 samples/s (57 samples/s for thermocouples)

Frequency Measurements

  • DataShuttle/USB 55: 2 frequency/pulse input channels
  • DataShuttle/USB 56: 4 frequency/pulse input channels
  • Operating Modes: Pulse count (totalize), frequency, and duty cycle
  • Frequency Response: DC to 1 MHz
  • Input Range: ±15V absolute maximum, <1.3V (low), >3.8V (high)
  • Pull-up Resistor: 27K Ohm to +5V for switch or relay sensing
  • Debouncing: None, 0.8, 3.2, or 13 msec (software selectable)
  • Totalize: Up to 2^32 counts/scan
  • Frequency Resolution: Up to 7 digits, user-selectable

Digital I/O

  • Each I/O line is individually programmable as input or output
  • Each I/O line includes an open-collector driver with a 27K Ohm pull-up resistor to +5V for output, and a Schmitt-trigger input buffer
  • DataShuttle/USB 55: 8 digital I/O lines
  • DataShuttle/USB 56: 16 digital I/O lines
  • Over-Voltage: +20V for up to 1 minute
    Output:
    • Voltage Range: 0 to +5V with no external pull-up resistor; 0 to +15V with external pull-up
    • Maximum Sink Current: 150 mA/output continuous, 500 mA output peak (<100 µs), 150 mA total continuous (per bank of 8 outputs)
    • Output Resistance: 10 Ohms max
    • Output Updates: Outputs may be changed
      via program control
  • Input:
    • Voltage Range: 0 to +15V
    • Thresholds: <1.3V (low), >3.8V (high)

DataShuttle

General

  • Power Requirements: +5 to +9 VDC @ 300 mA; AC adaptor included
  • Operating Temperature: 0° to 50°C
  • Dimensions: 190 mm x 140 mm x 50 mm (7.5” x 5.5” x 2.0”)
  • Resolution* Speed
  • 20 bits** 45/55 Hz†
  • 18 bits 45/55 Hz†
  • 16 bits 200 Hz
  • 14 bits up to 500Hz
  • 12 bits up to 500Hz
  • 11 bits up to 500Hz

* Resolution refers to full-scale readings; up to 10 times more resolution is provided at the lower 10% span of each range
** 20 bits of resolution are possible in the 18 bit resolution mode for inputs <10% of full scale
† 45 Hz when in 50 Hz low-noise mode, 55 Hz when in 60 Hz low-noise mode

Voltage Measurements

  • (Applies to General Purpose version, Thermocouple version, and when the range resistor is removed, the RTD version)
  • 8 differential inputs with screw terminal signal connections
  • Ranges:
    6 ranges, individually software selectable: -5 mV to +50
    mV, -25 mV to +25 mV, -50 mV to +500 mV, -250
    mV to +250 mV, -1V to +10V, -5V to +5V
  • Input Protection: 50V continuous, 150V for one second
  • Input Resistance: >109 Ohm (Ω)
  • Input Current: <3 nA
  • Accuracy: 0.05% of reading, +0.01% of range (15° to 35°C)
  • Temperature Coefficient: ±(40 ppm + 1 µV), 0° to 15°C, and 35° to 50°C
  • Common Mode Rejection Ratio (CMRR): >80 dB, ±5V and 10V ranges, >100 dB on all other ranges
  • Normal Mode Rejection Ratio (NMRR): 50 or 60 Hz, 40 dB min @ 18-bit resolution

Thermocouple Measurements

  • Thermocouple Types: E, J, K, T, B, R, S, C, D, G, N
    • Thermocouple Model
    • Isothermal Block insures accurate readings even under thermally dynamic conditions
    • Accuracy: 0.3°C (with environment @ 15° to 35°C)
    • Resolution: 0.01°C
    • General Purpose and RTD Model (with range resistor removed)
    • Accuracy: 0.8°C
    • Resolution: 0.01°C
    • Isolated 5B Model
    • Accuracy: 1.0°C

RTD Measurements

  • 8 channels of two- or three-wire measurements
  • RTD Types: 100 Ohm Platinum (can also accommodate 50 and 200 Ohm)
  • Range- 200° to +115°C -200° to +750°C
  • Resolution 0.005°C 0.01°C
  • Accuracy 0.8°C 1.0°C (with environment @ 15° to 35°C)

Digital I/O

  • 8 I/O lines, individually programmable as input or output
  • Each I/O line can be TTL-compatible, or open-collector
  • Optional modules can be installed on each channel, providing high-isolation and high-voltage switching and sensing on a per-channel basis
  • Input Voltage Range: 0 to +7V max
  • Input Voltage Logic Levels: TTL-compatible
  • Output Voltage Range: 0 to 30V max (with usersupplied pull-up resistors)
  • Output Current Sink: 50 mA/channel continuous, 400 mA max per 8 channels
  • Speed: Is the same as the selected analogue input measurement speed

Analogue Output (Models with –AO)

  • Two independent 12-bit programmable channels
  • Voltage Output Ranges (software selectable): 0 to +10V, 0 to +5V, 0 to +2V, -5V to +5V, -2.5V to +2.5V, -1V
    to +1V
  • Voltage Output Speed: Up to 1500 Hz (maximum is equal to maximum analogue input speed)
  • Current Output Range: 4 to 20 mA (5 to 15V compliance)
  • Speed: Is the same as the selected analogue input measurement speed
  • Accuracy: ±0.024% of full scale
  • Output Protection: -12V to +15V in volts mode, 0 to +15V in current mode



The Next Steps

• Ask us a question
• Buy DataShuttle Now
• View DataShuttle Pricing
• Request a Brochure
• Download a Brochure

Learn more about DataShuttle

Product Overview

DataShuttle 2000
DataShuttle 3000
DataShuttle USB
Accessories

Latest DataShuttle Information

New Features in
DataShuttle/2000

Trigger Modes
Specifications

DataShuttle Service & Support

Search the Knowledge Base
Technical support request

adept

Top of the Page

Popular Links: ChemDraw | ChemOffice | Data Acquisition | Data Analysis | Easy-PC | EndNote | Maple | MapleSim | Mathcad | MathType | Quality Analyst | Reference Manager | VisSim

Our Privacy and Terms and Conditions Statement
All Trademarks Recognised. Copyright © 2008, Adept Scientific.
Site designed and maintained by Adeptise

Adept Scientific | 7909 Charleston Ct. | Bethesda | MD 20817 | Tel: +1 800 724 8380